Introduction
The purpose of revision hip arthroplasty is to reconstruct the hip to reproduce, as closely as possible, the form and function of the native joint. The indications for revision include significant polyethylene wear, fractures of the components, component malposition, hip instability, severe thigh pain, excessive damage to the femoral Morse taper, and severe infection [].
In revision hip surgery, the removal of well-fixed cementless components can be extremely demanding, time consuming, and requiring patience and caution to limit the amount of host bone destruction. Many surgeons opt to retain a well-fixed acetabular cup unless it is malpositioned or shows signs of impingement or severe wear. Fortunately, indications for implant removal are scarce, and most of the contemporary cementless components perform very well through improved bony in-growth [].
Because of the diversity of the components and the methods used to secure them, an equal diversity of approaches and tools are necessary for component extraction. Component removal in total hip arthroplasty revision is a crucial and essential step in the operation because it dictates the possibilities for component reimplantation and reconstruction, and, therefore, of the patient outcome. It is imperative that the surgeon be comfortable and familiar with the basic techniques of implant removal. On the other hand, the implant industry has responded accurately to the needs of the revision surgeon and has developed instruments to overcome such issues with less effort [].
We describe a technique to remove a well-fixed polyethylene titanium-coated RM cup using acetabular reamers in a revision total hip arthroplasty. The reamers’ action reduces the thickness of the polyethylene to a thin lamina, and this way the cementless cup can be removed easily, minimizing acetabular bone loss. The cup was removed by a significant polyethylene wear at 20 years of follow-up.
Presentation
A 75-year-old man presented a bilateral hip arthroplasty. Clinical examination and radiographic findings showed aseptic loosening of a cementless prosthesis on the left side 20 years following the procedure.. The polyethylene titanium-coated RM acetabular cup was well positioned and well fixed; no radiolucent lines were present in the bone/implant interface. The isoelastic femoral stem showed a mechanical instability, the two screws of stem being broken (Figure ).
Management
Revision hip arthroplasty was performed with the patient in the lateral decubitus position under general anesthesia; the hip was exposed through a standard posterior approach. The femoral implant was loose in the femoral cavity and was removed without difficulty. Metallic debris was found in the soft tissues of the femoral implant/bone interface.
The acetabular component was a monobloc hemispherical cup manufactured from ultra high molecular weight polyethylene (UHMWPE), with a pure titanium particle-coated surface. With heat and pressure, the particles are blasted into the polyethylene surface. The coating promotes osseointegration. Stability of the cup is provided by two anchoring pegs on the weight-bearing part on its outer surface. The inclination of pegs and holes diverges by 5 degrees, providing a press-fit effect that increases the rigidity of the primary fixation. This is supplemented by screws inserted through the periphery of the cup [].
The RM cup was removed because significant wear of the polyethylene surface was presented on direct observation and by placing an appropriately sized head into the cup.
First, we removed the three screws of the cup. The polyethylene layer of the cup was sequentially reamed using powered acetabular reamers.
The first reamer was a small standard-powered cheese grater acetabular reamer (40 or 44 mm). It was placed in the middle of the inner part of the cup (Figure ). Reamers of increasing sizes were used successively to wear the polyethylene layer (Figure ). A minimal amount of polyethylene debris ("polyethylene chips") was removed from the articular soft tissues to avoid any biological reaction.
In this case, we used chisels to weaken the superior part of the implant and to create a roughened surface for the reamer action. When the polyethylene was thin enough, it was removed easily by hand tools without the risks of bone fractures and bone stock loss. This technique is not time-consuming; it took 12 minutes.
The cup surface was osseointegrated without interposition of soft tissues in the bone/implant interface by a process of contact osteogenesis.
A metallic support ring and a cemented polyethylene cup were implanted. The femur was reconstructed with a revision conical stem and morselized cancellous bone allograft, cryopreserved [11]. The bone allograft was impacted to fill the space between the proximal part of the implant and the cortical native bone (Figure ). No complications were reported in the perioperative course and during the hospitalization period.
Figures
Discussion
A major concern during revision hip arthroplasty is the acetabular bone loss during the extraction of well-fixed acetabular components. In fact, removing a well-fixed cementless cup is not always an easy task as it may lead to significant bone loss, particularly in the medial wall of the acetabulum, and subsequently result in a higher incidence of component loosening [12].
For safe component removal during revision hip procedures, it is essential to use the appropriate equipment and surgical techniques. During the last two decades, continuous efforts have resulted in the development of better tools to remove the cemented and cementless components. Hand tools such as osteotomes and chisels, ultrasonic devices, and ballistically driven chiseling systems controlled by an endoscope have been used with good results. There are no controlled studies comparing the outcomes of the different tools and techniques because often more than one device or approach is used simultaneously in the same revision procedure [].
Removal of an ingrown, fully porous-coated acetabular cup is a difficult procedure because there is no interface between host bone and the cup. The procedure begins with complete acetabular rim exposure and extraction of the polyethylene. Because of the problems encountered in trying to replace an acetabular component associated with bone loss, surgeons have begun retaining the well-fixed component in situ by exchanging the polyethylene liner while attempting to encourage bone formation through grafting the defects with bone or bone substitutes []. This is typically done in older patients as a last resort.
If screws are present, it is imperative to have the appropriate screwdrivers available during the cup removal. A high-speed metal burr can be used to cut off the screw head if the screw is stripped or if difficulty is incurred during attempted extraction. In these situations, we can use pliers, small trephines, or screw removal instruments to remove the remaining part of the screw.
After adequate exposure, the implant-bone interface is disrupted with the use of narrow curved osteotomes. Short curved osteotomes are used in a circumferential manner, with gradual progression to longer osteotomes. Specific instruments have been developed and used to achieve the goal of safe component removal with minimal bone loss. The Explant system by Zimmer has been gaining popularity. This instrument allows safe penetration at the metallic cup-bone interface [12]. However, to remove a well-fixed titanium-coated RM acetabular cup, the technique that we described is more efficient in preserving host bone.
In special situations, the metallic cup can be sectioned with a high-speed metal cutting burr. The sectioned component pieces are removed sequentially with minimal destruction of the underlying bone [].
Despite the technology available and the numerous techniques described, the key element in component removal depends on a careful preoperative planning, technical expertise, and patience. Stepwise, logical progression can result in successful and complete removal of all components, whereas impatience can result in catastrophic bone loss and fracture [].
In cases of well-fixed polyethylene titanium-coated RM acetabular component, we can also use the general procedures recommended for removal of the cementless cups. However, there is always the risk of bone loss and bone fracture because the cup is osseointegrated.
For that, we would recommend wearing the polyethylene layer with powered acetabular reamers. When the polyethylene is thin enough, the remaining cup is removed easily by hand tools, despite the osseointegration of the cup surface. The acetabular bone stock is preserved and the risks of fractures and bleeding are minimized.
We have used this technique in 10 cases with excellent results; no complications were noted. To our knowledge and on reviewing the literature, these principles were applied only in cemented cups [13].
Conclusion
Removal of a well-fixed cementless titanium-coated RM acetabular cup can be associated with bone loss, bone fracture, and bleeding when curved gouges, osteotomes, and chisels are used to disrupt the implant-bone interface. We have found a technique using acetabular reamers that allow cup thinning to be simple, reproducible, non-costly, non-timing consuming, and safe. The acetabular bone stock is preserved and the risks of complications are minimized. This technique can also be performed in the revision of a well-fixed all-polyethylene cemented cup.
Ethical Approval
The investigation process was developed in accordance with ethical research principles. The patient was informed that data concerning the case would be submitted for publication, and he consented.
Author Information
Fernando M.J. Judas, Rui F. Dias, and Francisco M. Lucas are from the University Clinic of Orthopaedics, FMUC – Faculty of Medicine, University of Coimbra, Coimbra, Portugal.
Source
Judas FMJ, Dias RF, Lucas, FM. A Technique to Remove a Well-Fixed Titanium-Coated RM Acetabular Cup on Revision Hip Arthroplasty. Journal of Orthopaedic Surgery and Research 2011, 6:31 doi:10.1186/1749-799X-6-31. http://www.josr-online.com/content/6/1/31; accessed 12/3/12.
© 2011 Judas et al; licensee BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
References
Attachments:
img_2012_article_1_revisiontha_fig1_200x408.JPG (image/jpeg)
img_2012_article_1_revisiontha_fig2_300x202.JPG (image/jpeg)
img_2012_article_1_revisiontha_fig3_300x224.JPG (image/jpeg)
img_2012_article_1_revisiontha_fig4_200x465.JPG (image/jpeg)
img_2012_article_1_revisiontha_fig1_200x408.JPG (image/jpeg)
img_2012_article_1_revisiontha_fig1_200x408.JPG (image/jpeg)
Figure 1.jpg (image/jpeg)