Findings:
Film .1 is an AP film of the left hip taken when the patient had no hip pain. Film .2 is an AP film and Film .3 is a groin lateral film taken at the time of the patient’s bitter complaint of buttock pain. The films show no evidence of loosening of either the cemented femoral or cementless acetabular component. The clinician insists that the patient must be having pain due to a low back syndrome. She insists that it is due to malfunction of her hip. Can you resolve the dilemma?
Category:
Adult Reconstruction
Region:
Hip
What is your diagnosis?
Click for diagnosis & discussion
Diagnosis:
Failed left total hip due to particle disease
Discussion:
This is a tough case. The AP films show no significant difference from one another. However, look carefully at the groin lateral film. This film is useful not only to evaluate retroversion versus anteversion of the acetabular component, but also to look for loosening and other complications of the prosthesis. Note that there is a large lytic lesion which is rather well-marginated involving a large part of the posterior fixation of the acetabulum (Film .3a, arrows). Since there is no gross loosening of either component, one might be tempted to suggest this could be due to metastatic disease or multiple myeloma since the patient certainly falls into the correct age group. However, look again at the AP film. Specifically, look at the centering of the femoral head in the acetabulum. Over the 18 months, the femoral head has migrated superiorly within the acetabulum (Film .4). The only way this can happen is through breakdown of the polyethylene cup which is placed within the metallic acetabulum backing. Given that this patient must have polyethylene breakdown, we can now deduce that the lytic lesion in the posterior acetabulum is due to particle disease.
Note that the prosthesis is a hybrid one, with cement being used to secure the femoral component, while the acetabular component is cementless. It has been observed that femoral components generally fail earlier than acetabular components, but that femoral components do better if they are cemented while long-term acetabular failure seems to be related to cement use. Therefore, it is felt that this hybrid method of securing the components may lengthen the useful life of a prosthesis in the majority of patients.
Reference:
Attachments:
422_1.JPG (image/pjpeg)
422_2.JPG (image/pjpeg)
422_3.JPG (image/pjpeg)
422_3A.JPG (image/pjpeg)
422_4.JPG (image/pjpeg)