Figures
Findings:
This is a cementless total knee arthroplasty. An effusion is present. There is superior subsidence of the patellar button. Fragmentation is seen involving the metallic portion of the patellar button. Several metallic fragments are seen lining the polyethylene portion of the component.
Category:
Adult Reconstruction
Region:
Knee
What is your diagnosis?
Click for diagnosis & discussion
Diagnosis:
Failed patellar component
Discussion:
This patellar component has failed in two ways. First of all, it shows evidence of chronic superior subsidence. Note the sclerotic line inferiorly in the patella as seen on the lateral film. This was the original placement of the button. Secondly, the metallic backing of the patellar component has actually fractured. Many small fragments are seen collecting on the polyethylene component (Film .2a). This of course makes the patient a set up for development of particle disease. Remember that particle disease can occur from cement, metal, or polyethylene particles if they are ground down to the critical size required for incitement of the inflammatory response.
Reference: