Which of the following factors will increase the rigidity of an external fixator?
1) decreased pin diameter
2) decreased pin number
3) increased pin spread within a segment
4) increased bone to rod distance
5) Increased distance between fragment pin sets
Figures
Answer (click for answer)
Preferred answer # 3
The most important factors that effect the rigidity of an external fixator is pin number, pin diameter, pin spread and distance between the connecting bar and the bone. By increasing the number of pins in the fracture fragments it will add rigidity to the ex fix. By increasing the diameter of the pins it strengthens the pin, which in turn will strengthen the ex, fix. By increasing the spread within a segment it adds rigidity to the ex fix i.e. by having one pin close to the fracture site which is unstable offset by having another pin as far from the site as possible adds the rigidity since it will be stabilized by the adjacent joint. Thus when reviewing the other choices the construct would be weakened by decreasing pin number or diameter . Increasing the bone to rod distance would create a higher profile construct, which would be less rigid.
Reference: Buckwalter, J.A. Einhorn, R.A. Simon, S.R.: Orthopaedic Basic Science: Biology and Biomechanics of the Musculoskeletal System, ed2. Rosemount, IL, American Academy of Orthopaedic Surgeons, 2000, pp372-379