Question
Following surgical treatment of an intertrochanteric femur fracture with a fixed angle dynamic hip screw, what factor has the greatest effect on the rate of implant cutout?
1- Length of the side plate
2- Length of the barrel
3- Angle of the device
4- Use of a compression screw
5- Tip-apex distance
Figures
Answer (click for answer)
Preferred Response: 5
Discussion:
Failure of fixation of peritrochanteric femur fractures treated with fixed-angle sliding hip screw devices is most closely related to the position of the lag screw in the femoral head. The sum of the distance from the tip of the lag screw to the apex of the femoral head as measured by both AP and lateral radiographs is called the tip-apex distance. In the study by Baumgaertner et al, it was shown that a tip-apex distance of 25mm or less resulted in no failures of fixation due to cut-out. They did however establish a relationship between an increasing tip-apex distance and the rate of cut-out. In the patients who were noted to have a failure of fixation secondary to implant cut-out the average tip-apex distance measured 38mm with a range of 28-48mm. Furthermore, for the patients whose tip-apex distance measured less than 30mm the rate of cut-out was less than 2% versus patients with a distance of greater than 30mm whose cut-out rate was 27%. At 35mm the rate was 36% and at 45mm or more the rate of cut-out was 60%.
Reference: Kellam JF, Fischer TJ, Tornetta P III, Bosse MJ, Harris MB (eds): Orthopaedic Knowledge Update: Trauma 2. Rosemont, IL, American Academy of Orthopaedic Surgeons, 2000, pp 125-131.