Year |
2002 |
|---|---|
|
Question Number |
243 |
|
Category |
Hip and Knee Reconstruction |
Question
The tibial insert shown in figure 68 was removed after 10 years in viv. Failure was most likely the result of
1) knee instability
2) infection
3) malalignment
4) third body abrasive wear
5) the sterilization method
Figures
Answer (click for answer)
Preferred Response: 5
This is a classic picture of poly delamination wear associated with gamma irradiation in an oxygen containing environment. This study linking wear with sterilezation lead to research yielding the current sterilization in a Nitrogen environment to prevent free radical generation and subsequent breakdown of crosslinking.
Reference: Collier JP, et al: Analysis of the failure of 122 polyethylene inserts from uncemented tibial knee components. Clinical Orthopedics 1991, 273:232-242.