Year |
2003 |
|---|---|
|
Question Number |
62 |
|
Category |
Hip and Knee Reconstruction |
Question
Early failure of a unicompartmental knee arthroplasty that is the result of polyethylene wear is primarily caused by
1- malalignment
2- instability
3- metal backing of the tibial component
4- gamma irradiation sterilization and shelf storage in air
5- obesity
Figures
Answer (click for answer)
Answer: 4
Free radicals created by gamma radiation in air make polyethylene prone to oxidation and decrease its mechanical toughness. The level of oxidation before implantation is directly related to the duration of exposure to oxygen before implantation (shelf life).
75 unicompartmental knee arthroplasties (62 pts) for medial compartment DJD.
73 (60 pts) had tibial bearings with > 4 yrs shelf life.
2 pts had tibial bearings with < 2 yrs shelf life wear were doing well.
37 knees (51%) in 32 pts required revision at a mean 18 mos (range 7 to 30) after arthroplasty.
The wear rate of poly increased as the shelf life increased.
4 to 5 yr shelf life: avg wear rate 0.9 mm/yr
5 to 6 yr shelf life: avg wear rate 1.6 mm/yr
to 7 yr shelf life: avg wear rate 1.7 mm/yr
Reference: McGoven et al. Rapid polyethylene failure of unicondylar tibial components sterilized with gamma irradiation. JBJS-A 2002; 34: 901-906.